The corrosion properties of multicomponent primarily casted and 100% recasted Ni- and Co-based dental alloys produced by the methods of powder metallurgy (PM) have been studied. Alloys are designed for porcelain fused to metal (PFM) reconstructions and have been investigated by means of two methods: static immersion in solutions which pH mimics aggressive environment in human cavity and by open circuit potential (OCP) measurements. The first one has been used to determine amounts of Ni2+ and Co2+ ions released under the influence of corrosion phenomena in standardized conditions. The OCP method has been used to assess the corrosion stability of primarily casted and 100% recasted Co- and Ni- alloys in some typical media: physiological solution, 0.05M HCl, Coca-Cola and fresh orange juice. Correlations between the composition of specimens, acidity of the environment and the corrosion resistance of the alloys have been established. It was shown that the amount of Ni2+ and Co2+ emitted during the immersion tests were far below the standardized threshold limiting application of dental alloys. It was also shown that all Co-base alloys have higher stability and the decreasing of pH leads to a decrease of the corrosion resistance. Results show that there was no significant difference in the corrosion potential values among the groups of primarily cast and 100% recast Ni- and Co-base metal alloys in HCl at pH 1.3. It was demonstrated that the high corrosion resistance and hence the strong biocompatibility of Ni- and Co- based dental alloys are result of the innovative PM production method used.
Key words: nickel and cobalt base dental alloys, corrosion, biocompatibility
Topic: CHEMISTRY